Dako Hybridizer | Handbook 25
When the fixed temp set temperature is reached the display prompts to “Add Slides and Close Lid”.
Before adding slides insert two Humidity Control Strips into the inside slide lid. After slides have been
added saturate strips with distilled water or equivalent (approx. 13 mL for dry strips) and close lid.
Cursor highlights “Start” line. Press “Enter” button again to run the program. (To return to the Main
Menu, move the cursor to highlight “Main Menu” line and press “Enter” button).
Display indicates present temperature of slides.
Timer counts elapsed time. (Pressing “Enter” button will reset timer to zero).
Use Arrow keys to move to “End PGM/Main Menu” line and press “Enter button to accept.
Abort Program in Process
To end a program in process press “STOP button, three beeps will sound. Use arrows to move cursor to
“Yes” line and press “Enter” button to accept. (Program will continue to run until “Yes” or “No” has
been accepted.)
Note: The Hybridizer prompts, “Are You Sure?” This measure is to prevent accidental disruption of a
program in process.
Fan will turn on and cool to 37°C if the slide temperature is above the default temperature of 37°C.